EBOV Soluble GP (sGP) ELISA kit is available at Gentaur for Next Week Delivery.
Product Type: kit
Host: N/A
Clonality: N/A
Applications: ELISA
1. Purpose:
For the quantitative measurement of EBOV soluble glycoprotein in
mouse and non-human primate sera
2. Reagents supplied:
3. Reagents required but not included in the kit:
- DPBS 1X, sterile (MediaTech/Corning cat# 21-031-CM) stored at
ambient temperature, for diluting coating antigen. - StartingBlock T20 (PBS) Blocking Buffer (Pierce/Thermo cat# 37539)
stored at 2-8C, for blocking and as diluent for standard, samples,
detection antibody, and tertiary antibody - DPBS powder (MediaTech/Corning cat# 55-031-PB) stored at 2-8C,
for preparing ELISA Wash Buffer - TWEEN-20 (Acros cat# 23336-0010), stored at ambient
temperature, for preparing ELISA Wash Buffer Deionized water
4. Materials required but not included in the kit:
- MaxiSorp flat bottom, polystyrene, 96-well plates (Nunc cat#
439454) - Polypropylene TiterTubes, maximum volume for each tube = 1 mL
(Bio-Rad cat# 223-9391) or equivalent, used to prepare standard
and sample dilutions - Microplate sealing film
- Polypropylene 15 mL and 50 mL conical tubes
- Reagent reservoirs
- Absorbent papers
5. Equipment required but not included in the kit:
- Automatic plate washer (example: BioTek Elx450)
- Plate reader with capability of measuring absorbance at 650 nm
(example: Molecular Devices plate reader) - Software for graphing the standard as a 4PL curve and for calculating
the unknown samples from the standard curve (example: Softmax
software) - Single-channel and multi-channel pipettes
6. Assay Procedure:
- Prepare Capture antibody solution
• Briefly spin the Capture Antibody vial and gently mix
by pipetting up and down
• Dilute 1:96 in DPBS 1X to target 4 µg/mL
• Example: For one full plate, add 116 µL Capture
antibody to 11 mL of DPBS 1X - Add 100 µL/well of Capture antibody solution to the MaxiSorp
plate. Cover plate using plate sealing film. Incubate covered
plate overnight at 2-8C. - The following day, equilibrate plate and StartingBlock Buffer to
ambient temperature for at least 15 min. - Empty contents from the plate and wash 3 times (each time 300
µL/well) with Wash Buffer using an automatic plate washer or
multi-channel pipette. - Add 200 µL/well of StartingBlock Buffer to block non-specific
binding. Incubate for at least 45 min at ambient temperature. - During blocking step, prepare dilutions of STANDARD and
UNKNOWN test samples in TiterTubes.